Braun Stimuplex HNS 11 HNS 11 Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Medição, teste e controlo Braun Stimuplex HNS 11 HNS 11. Braun Stimuplex HNS 11 HNS 11 User's Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 28
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 0
Stimuplex
®
HNS 11
Nerve Stimulator for
Regional Anaesthesia,
Regional Analgesia, Neurology
User manual
Vista de página 0
1 2 3 4 5 6 ... 27 28

Resumo do Conteúdo

Página 1 - Stimuplex

Stimuplex® HNS 11Nerve Stimulator forRegional Anaesthesia,Regional Analgesia, NeurologyUser manual

Página 2

Pulse LED displayThe small orange LED above the LCD display flashes according tothe pulse frequency set (1 Hz: once per second; 2 Hz: twice persecond)

Página 3 - Contents Page

The toneEach generated pulse produces a tick. When a stimulation currentflows to the electrodes (greater than 2% of the maximum stimula-tion adjustabl

Página 4 - Checks before start-up

Special technical featuresThe configuration of the electrode connection provides protectionagainst a wrong polarity connection of the electrode cable.

Página 5

BatteryThe battery must be regularly checked for correct charge by press-ing the ”BAT” button. While the button is pressed, the existingbattery voltag

Página 6 - Technical description

Cleaning and disinfecting the Stimuplex® HNS 11Use only soft, damp cloths to clean and disinfect the Stimuplex®HNS 11 and the electrode cable. Water,

Página 7

For usage of HNS 11 please refer to the individual local lawrestricts (eg. maintaining a logbook, etc.).In case of conducting safety checks on the dev

Página 8

CautionElectromagnetic fields may interfere with the function of the nervestimulator. Do not use the stimulator in the proximity of equip-ment which p

Página 9 - The LED displays

Safety informationThe nerve stimulator should not be used on patients withimplanted electrical devices (e.g. cardiac pacemakers) withoutprior consulta

Página 10 - The LCD display

Instruments and accessoriesThe Stimuplex® HNS 11 nerve stimulator should only be operatedwith the accessories and instruments mentioned on page 13, 14

Página 11 - The tone

Product discription Canula Code number Sales unitØ x lengthStimuplex® A needles, 30° bevelA 25; 24 G x 1" 0.55 x 25 mm 4894251 25A 25; 22 G x

Página 12

Distributor:B. Braun Melsungen AGP.O. Box 11 20D-34209 MelsungenTel. (0 56 61) 71-0Fax (0 56 61) 71-3811Manufacturer:Stockert GmbHBötzinger Straße 72D

Página 13 - Warning!

Peripheral electrical nerve stimulationPrinciplesBy simplifying the accurate location of peripheral nerves, periph-eral electrical nerve stimulation (

Página 14 - Caution!

early detection of technical faults which could endanger thepatient and place the success of the anaesthesia in jeopardy.Performing PNS with the Stimu

Página 15

If the two values coincide, the total resistance is acceptable(impedance test) and the electric circuit is satisfactory. If theactual value is less th

Página 16

Literature23Literature1. Moore DC (1965), Regional block. A handbook for use in theclinical practice of medicine and surgery. Thomas, SpringfieldII. (

Página 17

Appendix AStimulation current display in the mA rangePatient current Measurement display Controller display0.00 mA 0.00 mA 0.00 mA0.05 mA 0.05 mA ± 0.

Página 18 - needles

Appendix C25Stimuplex® HNS 11 Nerve StimulatorMax. stimulation current as a function of tissue resistanceMax. stimulationcurrent [mA]Battery voltage [

Página 19

Stimuplex® D and Contiplex® D needles (DP: DE 3919666)Only 0.2 or 0.05 mAstimulation current requiredblunt30°pointed15°Stimulationcurrent[mA]Threshold

Página 21 - or Contiplex

B. Braun Melsungen AGP.O. Box 11 20D-34209 MelsungenTel (0 56 61) 71-0Fax (0 56 61) 71-3811www.bbraun.comW. 01.08.01/7 Nr. 604 3895hospital careCautio

Página 22

Contents PageChecks before start-up...

Página 23 - Literature

Checks before start-upThe Stimuplex® HNS 11 is a class lla medical device according toCouncil Directive 93/42/EEC. An item of medical equipment mayonl

Página 24 - Appendix B

In the operating mode, the LCD display shows the preselectedstimulation current set on the controller between 0.00 andapprox. 1.00 mA (in the 1 mA r

Página 25 - Appendix C

Technical description6. Refer to the Safety information section before using the Stimu-plex® HNS 11 on a patient.Any instrument behaving abnormally mu

Página 26 - Appendix D

Operating the nerve stimulatorSwitching on / switching offTurn the control knob clockwise out of the O position to switch onthe Stimuplex® HNS 11. A b

Página 27

Selecting the current rangeThe maximum current available can be selected by pressing the”mA (max)” button. The Stimuplex® HNS 11 nerve stimulator offe

Página 28 - Caution:

The LED displaysAll LEDs directly assigned to the buttons are used to monitor theselected operating mode. Since LEDs require relatively high power,the